IKA S 25 EC - T - C - 18 G Dispersing Tool, Ø 18 mm, with Temperature Sensor, 0020108998
€1,425.00
% Manufacturer's RRP: €1,900.00 25% below manufacturer's RRP
ⓘ This product description has been translated from German using AI. Despite careful review, we do not guarantee that the translation is error-free; in case of doubt, the German version shall prevail.
Product information "IKA S 25 EC - T - C - 18 G Dispersing Tool, Ø 18 mm, with Temperature Sensor, 0020108998"
The temperature sensor sits in the tool itself: the IKA S 25 EC - T - C - 18 G measures the temperature in the batch during dispersion, within a range of 0 to 125 °C, with an accuracy of plus or minus 1 Kelvin. The reading is shown on the drive; the sensor function can only be activated on the T 25 easy clean control.
Operating time and maintenance
The associated T 25 easy clean digital or control disperser records the operating time of the tool. When maintenance is required, a message appears on the display.
Biochemistry and pharmaceuticals
IKA particularly recommends this autoclavable tool for dispersing applications in biochemistry and pharmaceuticals, such as the emulsification of heat-sensitive active ingredient intermediates. Cleaning follows the ‘easy-clean’ principle: turn the clamp and lower it, pull out the shaft and rotor, and rinse the shaft tube from the inside.
Dimensions and resistance
Stator 18 millimetres, rotor 12.7 millimetres, gap 0.3 millimetres, working volume 10 millilitres to 1.5 l. Immersion depth 45 to 150 millimetres, installation length 191 millimetres. The final particle size is 10 to 50 micrometres for suspensions and 1 to 10 micrometres for emulsions. Materials in contact with the product are AISI 316L and ceramic; the maximum operating temperature is 120 °C; sterilisation is permitted using all methods except hot air. The tool weighs 0.634 kilograms.
Typical applications
- Dispersing and emulsifying with in-batch temperature measurement
- Emulsification of heat-sensitive active ingredient intermediates
- Dispersion applications in biochemistry and pharmaceuticals
- Disintegration of particles in solution
- Batches containing solvents or abrasive components
Technical details
| Stator diameter | 18 mm |
|---|---|
| Rotor diameter | 12.7 mm |
| Stator/rotor gap width | 0.3 mm |
| Processable volume (H2O) | 0.01 to 1.5 l |
| Maximum permissible speed | 25,000 rpm |
| Maximum peripheral speed | 16.6 m/s |
| Immersion depth | 45 to 150 mm |
| Installation length | 191 mm |
| Material in contact with the product | AISI 316L, ceramic |
| Temperature measurement range | 0 to 125 °C |
| Temperature measurement accuracy | ±1 K |
| pH range | 2 to 13 |
| Suitable for solvents | Yes |
| Suitable for abrasive substances | Yes |
| Max. operating temperature | 120 °C |
| Sterilisability | all methods except hot air |
| Final particle size of suspensions | 10 to 50 µm |
| Final particle size of emulsions | 1 to 10 µm |
| Weight | 0.634 kg |
Scope of delivery
- 1 x IKA S 25 EC - T - C - 18 G dispersing tool (single tool)
- The drive is not included in the Scope of delivery.
Without sensor, with sawtooth blades
The same size is available without a temperature sensor as the S 25 EC - C - 18 G. There is also the S 25 EC - T - C - 18 G ST, which features sawtooth blades for fibrous materials and a narrower gap. The 25-millimetre version with a sensor is the S 25 EC - T - C - 25 G.
Selection guide comparison
| Tool | Stator / Rotor | Working range (H2O) |
|---|---|---|
| S 25 EC - T - C - 18 G (this product) | 18 / 12.7 mm | 0.01 to 1.5 l |
| S 25 EC - C - 18 G | 18 / 12.7 mm | 0.01 to 1.5 l |
| S 25 EC - T - C - 18 G ST | 18 / 13.4 mm | 0.01 to 1.5 l |
| S 25 EC - T - C - 25 G | 25 / 17 mm | 0.05 to 2 l |
Individual requirements and procurement service